As can be seen from the chromatogram, deionized water in the 8.3min and 11.5min detection of two substances, of which 11.5min chromatographic peaks and Cl-retention time consistent, indicating that there is a trace of deionized water Cl- (signal to noise ratio For 4.6). F- (retention time is 6.2 min) is not detected.
Use a vial to prepare the various solutions before using deionized water for repeated cleaning and soaking for 4 hours. After soaking for 4 hours, the blank value of the solution was determined and the chromatogram was as follows.
As can be seen from the chromatogram, deionized water in the immersion capacity bottle 4 hours later, there are three substances were detected. The signal-to-noise ratio of Cl- in this sample is 5.4, and F- is not detected. After measuring the blank of deionized water and volumetric flask, a solution containing 2.5 mg / LF-, Cl- was injected into an ion chromatograph to give the following chromatogram.
As can be seen from the chromatogram, F-, Cl- are clearly detected in the chromatogram, in addition to three unknown components. The signal-to-noise ratios of F- and Cl- are 3.4 and 4.0, respectively (Cl-not blank).
A solution containing 5.0 mg / LF-, Cl- was injected into an ion chromatograph to give the following chromatograms.
As can be seen from the chromatogram, F-, Cl- are clearly detected in the chromatogram, in addition to two unknown components. The signal-to-noise ratios of F- and Cl- are 6.0 and 8.0, respectively (Cl-not blank). Conclusion The use of domestic PIC ion chromatography and five-pole conductivity detector, can be sensitive to the determination of mg / LF- and Cl-. Deionized water and volumetric flask blank samples show that F-in the chromatographic conditions there is no interference, can be sensitive and accurate determination. There is a trace of Cl- in the blank sample, and the signal-to-noise ratio is greater than three, so that several mg / LCl- in the actual sample can only be sensitive and can not be accurately determined. The use of more pure water as a solvent, when the deionized water and volumetric bottles and other blank samples Cl - less than three times the signal to noise ratio can accurately determine the trace of Cl-.